Completed tests:
1. Changed oil, rebuild plumbing, wired up extension cord and changed plug, demonstrated basic function.
2. Checked base vacuum: 1*10^-2Torr within ~ a minute and stable over ~5 hours. For comparison, Andrea thesis (pg 139) reports 0.3 Torr ultimate pressure while pumping on the empty column.
3. Plumbed up a flow meter in series with a needle valve. Air flow was adjusted from 8-18 CFH (0.13-0.30 CFM), calibrated for CO2, with corresponding input pressures 0.25 – 1.1 Torr. Andrea estimates a purge flow of 0.1cfm for the exo200 filter. For comparison, Andrea’s thesis (pg 141) shows tank pressures of 4-16 Torr over a purge flow range of 1-4 l/s (2-8 cfm), with the columns loaded.
4. System ran for 7.5hrs on 12/21/12 and for 26hrs on 1/2/13 with no gas load to check for stability. Base pressure was stable at 1*10^-2 Torr and no “hot electronics” smell in either case.
5. System ran from 1/3 to 1/7 with about 10cfh (as calibrated for CO2) of air. Base pressure stable around 0.2 Torr. No excessive heat or noise, some smell of pump oil.
6. Pumpdown tests on column 1/9. Column was blanked off on one side, connected to ISO80 valve input on the other with (all) the 3″ plastic tubing. Test sequence (repeated 3 times) was:
- Column full, valve closed.
- Pump started, pressure noted
- Valve opened, pressure noted
- Pump stopped, system vented
- Valve closed, oil level noted.
Plot of pressure vs time while pumping out one column below. The system reaches 2*10^-2 Torr in about a minute.
The oil level for the backing pump at the end of each trial (pump off) varied between the “off” and “on” marks that Andrea put on the sight glass. No trend observed.
